Mechanical Aortic Valves

Top Hat Aortic Options Tabulated to Patient's Annulus Diameter
Patient's Annulus Diameter(mm) 17 19 21 23
Applicable Top Hat Valve Sizes(mm) 19 21 21 23 23 25 25 27
Mean Pressure Gradients (mmHg) 19.4 12.3 12.3 9.8 9.8 8.3 8.3 8.8
Mean Velocity (m/s) 2.01 1.60 1.60 1.44 1.44 1.33 1.33 1.35
Peak Pressure Gradient (mmHg) 40.1 28.2 28.2 20.0 20.0 16.0 16.0 17.7
Peak Velocity (m/s) 3.12 2.48 2.48 2.15 2.15 2.02 2.02 2.01
Effective Orifice Areas (cm2) 1.0 1.5 1.5 1.6 1.6 2.0 2.0 2.4


R-Series and Standard Intra-Annular Aortic Performance
Valve Size(mm) 19 21 23 25 27 29 31
Mean Pressure Gradient(mmHg) 19.4 12.3 9.8 8.3 8.8 5.8 5.8
Mean Velocity (m/s) 2.01 1.60 1.44 1.33 1.35 1.10 1.10
Peak Pressure Gradient (mmHg) 40.1 28.2 20.0 16.0 17.7 10.8 10.8
Peak Velocity (m/s) 3.12 2.48 2.15 2.02 2.01 1.90 1.90
Effective Orifice Areas (cm2) 1.0 1.5 1.6 2.0 2.4 2.6 2.6


Mechanical Mitral Valves

Standard Mitral, Optiform Mitral, and Orbis Mitral Performance
Valve Size (mm) 23 25 27 29 31 33
Mean Pressure Gradient (mmHg) 7.0 4.1 3.9 3.3 3.3 3.4
Peak Pressure Gradient (mmHg) 18.0 9.4 10.5 10.1 8.6 7.6
Peak Velocity (m/s) 2.10 1.48 1.55 1.55 1.41 1.35
Pressure Half-time (ms) 104 81 78 67 83 79
Effective Orifice Area (cm2) 1.3 2.2 2.1 2.1 1.9 2.3
 
 